Fluke 96000 Series Service Manual page 109

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Instrument
10 MHz Output from
Frequency Standard
Figure 3-18. Equipment Connections - Microwave Output Frequency Accuracy Test Part A
4. Set the UUT as follows:
Ref Frequency Input
Mode
Signal
Ref Frequency Input
Frequency
Level
Output
5. Set the frequency counter as follows:
Function
Slope
Coupling
Input Impedance
Attenuation
Trigger Mode
6. Set the output frequency of the UUT to each of the frequencies listed in part
A of Table 3-28. At each frequency allow the frequency counter to take
several readings and settle; the settled reading must be within the tolerance
shown. At 1 GHz connect the UUT Microwave output to the counter channel
C input and select the counter Freq C function.
7. Set the UUT output to STBY. Remove all connections to the frequency
counter.
Disabled
Leveled Sine
Microwave Output
Disabled
10 kHz
+13 dBm
OPER
Freq A
Positive
AC
50 Ω
1x
Auto
Performance Tests
Performance Test
Frequency Counter
MHz
Front
!
!
!
Rear
External
Reference
Frequency
Input
3
hpn391.eps
3-67