Fluke 96000 Series Service Manual page 110

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

96000 Series
Service Manual
Part
A
B
8. Set the UUT output to STBY. Remove all connections to the frequency
9. Connect the 10 MHz output from the frequency standard to the EXT REF IN
10. Connect the UUT Microwave output to the input of the spectrum analyzer.
3-68
Table 3-28. Frequency Accuracy Test – Microwave Output
UUT Frequency
10 MHz
1 GHz
5 GHz
6.036 GHz
8.048 GHz
9.52 GHz
10 GHz
11.32 GHz
12.072 GHz
13.5 GHz
15 GHz
16 GHz
18.108 GHz
19.04 GHz
20 GHz
21.84 GHz
22.64 GHz
24 GHz
26.5 GHz
counter.
on the spectrum analyzer (rear panel) using a BNC male – male cable
assembly. Set the spectrum analyzer for external frequency reference.
See Figure 3-19 for equipment connections.
Tolerance
9.999 999 500 – 10.000 000 500 MHz
0.999 999 950 0 – 1.000 000 050 0 MHz
4.999 999 750 0 – 5.000 000 250 0 MHz
6.035 999 698 2 – 6.036 000 301 8 MHz
8.047 999 597 6 – 8.048 000 402 4 MHz
9.519 999 524 0 – 9.520 000 476 0 MHz
9.999 999 500 0 – 10.000 000 500 0 GHz
11.319 999 434 0 – 11.320 000 566 0 GHz
12.071 999 396 4 – 12.072 00 603 6 GHz
13.499 999 325 0 – 13.500 000 675 0 GHz
14.999 999 250 0 – 15.000 000 750 0 GHz
15.999 999 200 0 – 16.000 000 800 0 GHz
18.107 999 094 6 – 18.108 000 905 4 GHz
19.039 999 048 0 – 18.108 000 905 4 GHz
19.999 999 000 0 – 20.000 001 000 0 GHz
13.499 999 325 0 – 21.800 001 090 0 GHz
21.799 998 910 0 – 13.500 000 675 0 GHz
23.999 998 800 0 – 24.000 001 200 0 GHz
26.499 998 675 0 – 26.500 001 325 0 GHz