Fluke 96000 Series Service Manual page 115

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Level
Frequency
20 kHz
2.74 MHz
5.5 MHz
11 MHz
22 MHz
50 MHz
88 MHz
125 MHz
250 MHz
354 MHz
500 MHz
+16 dBm
1 GHz
1.25 GHz
1.4 GHz
1.8 GHz
2 GHz
2.7 GHz
5 GHz
9 GHz
10 GHz
12 GHz
13 GHz
18. Set the Spectrum Analyzer as follows:
PRESET
COUPLING
REF LVL
EXT REF
FREQ
SPAN
Table 3-29. Harmonics Test, Microwave Output
nd
2
Limit
Harmonic
40 kHz
-60 dBc
5.48 MHz
-60 dBc
11 MHz
-60 dBc
22 MHz
-60 dBc
44 MHz
-60 dBc
100 MHz
-60dBc
176 MHz
-60 dBc
250 MHz
-60 dBc
250 MHz
-60 dBc
708 MHz
-60 dBc
1 GHz
-60 dBc
2 GHz
-60 dBc
2.5 GHz
-55 dBc
2.8 GHz
-55 dBc
2.8 GHz
-55 dBc
4 GHz
-55 dBc
5.4 GHz
-55 dBc
10 GHz
-55 dBc
18 GHz
-55 dBc
20 GHz
-55 dBc
24 GHz
-55 dBc
26 GHz
-55 dBc
DC
+10 dBm
On
4.399994 GHz
2 MHz
rd
3
Harmonic
60 kHz
-60 dBc
8.22 MHz
-60 dBc
16.5 MHz
-60 dBc
33 MHz
-60 dBc
66 MHz
-60 dBc
150 MHz
-60dBc
264 MHz
-60 dBc
375 MHz
-60 dBc
750 MHz
-60 dBc
1.062 GHz
-60 dBc
1.5 GHz
-60 dBc
3 GHz
-60 dBc
3.75 GHz
-55 dBc
4.2 GHz
-55 dBc
5.4 GHz
-55 dBc
8 GHz
-55 dBc
8.1 GHz
-55 dBc
15 GHz
-55 dBc
-
-
-
-
-
-
-
-
Performance Tests
Performance Test
Limit
3
3-73

Advertisement

Table of Contents
loading

This manual is also suitable for:

96040a96270a

Table of Contents