32. Return the UUT output to -81dBm, set the UUT frequency to the next
33. Set the UUT to 10 MHz at -91dBm. Repeat steps 39 through 43 for the points
34. Set the UUT to STBY.
Table 3-21. 75 Ω Leveling Head Optional Ultra-Low Level Accuracy Test (96040A and 96270A)
Part
A
B
3-48
frequency listed in part A of Table 3-21, and repeat steps 29 through 32.
listed in part B of Table 3-21 using the calculation P
Frequency
10 MHz
100 MHz
300 MHz
1.4 GHz
2 GHz
10 MHz
100 MHz
300 MHz
1.4 GHz
2 GHz
Amplitude
Reading
-81 dBm
Reference
-101dBm
P
rel
-121dBm
P
rel
-81 dBm
Reference
-101dBm
P
rel
-121dBm
P
rel
-81 dBm
Reference
-101dBm
P
rel
-121dBm
P
rel
-81 dBm
Reference
-101dBm
P
rel
-121dBm
P
rel
-81 dBm
Reference
-101dBm
P
rel
-121dBm
P
rel
-91 dBm
Reference
-111 dBm
P
rel
-91 dBm
Reference
-111 dBm
P
rel
-91 dBm
Reference
-111 dBm
P
rel
-91 dBm
Reference
-111 dBm
P
rel
-91 dBm
Reference
-111 dBm
P
rel
= P
+ P
in step 30.
out
-91
rel
P
Tolerance
out
Reference
±1.50 dB
±1.50 dB
Reference
±1.50 dB
±1.50 dB
Reference
±1.50 dB
±1.50 dB
Reference
±1.50 dB
±1.50 dB
Reference
±1.50 dB
±1.50 dB
Reference
±1.50 dB
Reference
±1.50 dB
Reference
±1.50 dB
Reference
±1.50 dB
Reference
±1.50 dB