Fluke 96000 Series Service Manual page 102

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

96000 Series
Service Manual
The recommended Power Meter has Power Sensor calibration
factors stored internally in the sensor. Setting the input frequency in
the Power Meter automatically selects the appropriate cal factor for
the measurement. However, the Power Sensor may also be
accompanied by a table of calibration factors which must be applied
externally.
11. Allow the Power Meter reading to settle. Record the measured level in dBm
The Power Meter autorange boundary may coincide with a
measurement point. To avoid unwanted Power Meter autoranging,
manually lock the Power Meter range once the 100 kHz reading is
obtained and maintain the Power Meter range lock for all
subsequent frequencies at this amplitude.
12. Set the UUT frequency to the first test frequency listed in Table 3-25.
13. Enter the measurement frequency into the Power Meter and allow the Power
14. Check that the value of P
If a Power Sensor with other than the recommended 2.92 mm RF
input connector is used, the insertion loss of any connector adapter
must be taken into account at each measurement frequency.
15. Set the UUT to the next frequency point at this amplitude listed in Table 3-25,
The highest frequency tested is 26.4 GHz. This frequency helps
prevent attempted measurements at the 26.5 GHz maximum
frequency limit of the Spectrum Analyzer/Measuring Receiver used
in subsequent parts of this procedure. This small frequency margin
allows for any residual frequency difference errors that may
otherwise take the Spectrum Analyzer/Measuring Receiver beyond
its operating frequency range.
16. Set the UUT to 100 kHz at the next amplitude listed in Table 3-25, and repeat
17. Set the UUT output to STBY. This concludes testing for UUTs not fitted with
3-60
as P
for use later in this procedure.
2
Meter reading to settle. Record the measured level in dBm as P
the UUT output P
= P
out
in Table 3-25.
and repeat steps 13 through 15.
steps 11 through 15. For instruments fitted with the Low-level Microwave
Output option, when the test amplitude is -7 dBm at each frequency, record
the value of P
at -7 dBm as P
out
procedure.
the Low-level Microwave Output option.
Note
Note
+ (P
– P
).
1
3
2
is consistent with the typical performance shown
out
Note
Note
. These values will be used later in this
-7
. Calculate
3

Advertisement

Table of Contents
loading

This manual is also suitable for:

96040a96270a

Table of Contents