Fluke 96000 Series Service Manual page 103

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Table 3-25. Microwave Output Level Accuracy Test, High Frequency Test Points
Low Level Microwave
Output Option?
All UUT configurations
Non-Low Level MW O/P
Option only
Low Level Microwave
Output Option only
Amplitude
Frequency
100 kHz
10 MHz
100 MHz
1 GHz
+13 dBm
4.024 GHz
10 GHz
15 GHz
20 GHz
26.4 GHz
100 kHz
10 MHz
100 MHz
1 GHz
+3 dBm
4.024 GHz
10 GHz
15 GHz
20 GHz
26.4 GHz
100 kHz
10 MHz
100 MHz
1 GHz
-4 dBm
4.024 GHz
10 GHz
15 GHz
20 GHz
26.4 GHz
100 kHz
10 MHz
100 MHz
1 GHz
-7 dBm
4.024 GHz
10 GHz
15 GHz
20 GHz
26.4 GHz
Reading
P
Typical Tolerance
out
P
Reference
2
P
±0.5 dB
3
P
±0.5 dB
3
P
±0.5 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
Reference
2
P
±0.5 dB
3
P
±0.5 dB
3
P
±0.5 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
Reference
2
P
±0.5 dB
3
P
±0.5 dB
3
P
±0.5 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
±1.0 dB
3
P
Reference
2
P
±2.0 dB
3
P
±2.0 dB
3
P
±2.0 dB
3
P
±2.0 dB
3
P
±2.0 dB
3
P
±2.0 dB
3
P
±2.0 dB
3
P
±2.0 dB
3
3
Performance Tests
Performance Test
3-61