Fluke 96000 Series Service Manual page 177

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Level Accuracy - 50
Use an AC Measurement Standard, a power meter and power sensors, and a
high performance spectrum analyzer/measuring receiver to make the
measurements shown in Tables A-16 and A-17.
Level
+16 dBm
+13 dBm
+3 dBm
Ω
Levelling Head (96040A and 96270A)
Table A-12. Level Accuracy (50 Ω) Test
Frequency
1 kHz
20 kHz
100 kHz
10 MHz
100 MHz
300 MHz
1.4 GHz
1 kHz
20 kHz
100 kHz
10 MHz
100 MHz
300 MHz
1.4 GHz
3 GHz
4 GHz
1 kHz
20 kHz
100 kHz
10 MHz
100 MHz
300 MHz
1.4 GHz
3 GHz
4 GHz
Performance Test Record
Performance Test Record
Actual
Tolerance
±0.03 dB
±0.03 dB
±0.03 dB
±0.05 dB
±0.05 dB
±0.07 dB
±0.2 dB
±0.03 dB
±0.03 dB
±0.03 dB
±0.05 dB
±0.05 dB
±0.07 dB
±0.2 dB
±0.3 dB
±0.3 dB
±0.03 dB
±0.03 dB
±0.03 dB
±0.05 dB
±0.05 dB
±0.07 dB
±0.2 dB
±0.3 dB
±0.3 dB
A