Fluke 96000 Series Service Manual page 113

Rf reference source
Hide thumbs Also See for 96000 Series:
Table of Contents

Advertisement

Signal content at the required harmonic frequency may be well
below the tolerance limit, and difficult to observe on the Analyzer
display with the above settings. Provided the harmonic signal is
below the allowed limit, the test result may be recorded as "Pass".
However, if it is desired to measure the actual signal level, reduce
the analyzer resolution bandwidth setting and/or reduce the analyzer
reference level setting. Remember to restore the analyzer reference
level setting to that shown in step 4 before changing the UUT to the
next fundamental output frequency, otherwise the Analyzer may be
overloaded and invalid results obtained.
9. Set the Spectrum Analyzer as follows:
FREQ
60 kHz (3
MARKER → PEAK
10. Using the Spectrum Analyzer marker functions, measure the marker
amplitude on the Spectrum Analyzer for the UUT's 3
displayed (delta marker) value must be within the limit listed in Table 3-29 for
rd
3
harmonic.
Instrument
Figure 3-20. Equipment Connections – Microwave Output Harmonics and Spurious Content Test
Note
rd
harmonic)
Front
Rear
Performance Tests
Performance Test
rd
harmonic. The
Spectrum Analyzer
Ext Ref Input
3
RF Input
hpn372.eps
3-71