Running Multiple-Vial Analyses

Single-Level Calibration Sequences

Single-Level Calibration Sequences

The following procedure describes how to set up a calibration sequence for an
analysis which uses single- level calibration.
There is one calibration standard (C) and 9 samples (S).
The analysis requires that:
The example assumes that the sequence table is empty. If the sequence table is
not empty, use the Delete button to delete all lines or use File - New.
1 Select the Sequence screen.
2 Move to line 1 and press Insert.
3 Select Method and then move to METH1 and press Load.
4 Move to the End of sequence list, select the Wizard and enter:
Vial Range
#Inj.
5 Select Calibration to display the Calibration Settings screen.
G4208A User's Guide
each sample is analyzed in duplicate,
the calibration standard is analyzed once before the samples and re- analyzed
once after every 2 samples,
C SS C SS C SS C SS C S C
the calibration standard is located in position 90 and the 9 sample vials are in
positions 1 to 9 of the autosampler tray, and
the method called METH1 is used for the samples and standards.
1 – 9
2
120