Agilent Technologies InfinityLab LC Series User Manual page 147

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

5
Maintenance and Repair
Firmware Updates
Upgrade from A.05.1x to B.xx
1 When the Agilent 1100/1200 series modules are on a firmware platform
2 Connect the Instant Pilot (with A.05.1x firmware) to one of the modules (if not
3 After boot, enter the Maintenance screen and press the Controller button.
4 Press the Transfer button. The Instant Pilot will switch into resident mode
5 Select the Instant Pilot for update to the target firmware revision
6 Start the update.
7 When finished, the Instant Pilot boots in main mode and shows all modules of
8 Close the connection and the LAN/RS- 232 Update Tool 2.10.
G4208A User's Guide
A.05.09/1x, then update the modules to platform A.06.xx
already done).
Figure 101 Update Information for firmware A.05.11
(black screen) and the Update Tool will list the Instant Pilot.
the system in the Welcome screen.