5
Maintenance and Repair
Firmware Updates
Upgrade from A.05.11 to A.05.1x
If the Instant Pilot needs to be updated with a new firmware revision , follow
these steps:
1 Disconnect all 1100/1200 series module from that module that hosts the
2 Upgrade this module to firmware A.06.xx first (otherwise the Instant Pilot is
3 After boot, enter the Maintenance screen and press the Controller button.
4 Press the Transfer button.
5 Select the Instant Pilot for update to the target firmware revision
6 Start the update.
7 After boot of the Instant Pilot,
8 Press Cancel (red cross button) on the LAN/RS- 232 Update Tool 2.10 to stop
9 Downgrade the 1100/1200 module to the target firmware (A.05.11/13 or
10 When finished,
11 Close the connection and the LAN/RS- 232 Update Tool 2.10.
G4208A User's Guide
LAN/RS- 232 interface.
not visible in the LAN/RS- 232 Update Tool 2.10).
The Instant Pilot will switch into resident mode (black screen) and the Update
Tool lists now the Instant Pilot.
the Instant Pilot shows "Scanning System ..."
the LAN/RS- 232 Update Tool 2.10 indicates "100% updated" but does not
display the result dialog.
the Update process.
The Instant Pilot is no longer listed in the tree view of the Update Tool.
A.05.09/10).
the LAN/RS- 232 Update Tool 2.10 shows all 1100/1200 modules without
the Instant Pilot.
the Instant Pilot shows all modules of the system in the Welcome screen.
148