Working with the Instant Pilot
Simultaneous Execution with Software
A method that is available on the Instant Pilot as a protected method can be
NOTE
modified by the Agilent ChemStation and then saved on the Instant Pilot without
any warning.
The method will be changed in the instruments, but the file cannot be saved
anyway.
G4208A User's Guide
accessible from the Agilent ChemStation). To have a method available on
both controllers, proceed as follows:
If the method is on the Agilent ChemStation and should be saved on the
Instant Pilot or USB Flash Drive, load the method on Agilent ChemStation and
then save the method on the Instant Pilot (or USB Flash Drive) with Method —
Save As.
If the method is on the Instant Pilot or USB Flash Drive and should be saved
on the Agilent ChemStation, first load method DEF_LC.M on the Agilent
ChemStation (to have no additional parameter in the format) and then load
the required method on Instant Pilot. Then save the method on the Agilent
ChemStation with the same name.