Running Multiple-Vial Analyses
Re-calibrating With Multiple Groups of Standards
4 Move to the End of sequence list, select the Wizard and enter:
Vial Range
#Inj.
5 Select Calibration to display the Calibration Settings screen.
6 Change the settings according to
7 You may press Preview to view the result.
8 Select Done to accept entries.
9 Move to the End of sequence list and press Insert.
Vial Range
#Inj.
10 Select Method and then move to METH2 and press Load.
You have to insert METH2/METH3 before/after each calibration sample line(s).
NOTE
Use copy/paste or use insert/parameter/stoptime and only one method.
11 Move to the End of sequence list, select the Wizard and enter:
12 Select Calibration to display the Calibration Settings screen.
G4208A User's Guide
7 to 12
1
Figure 84
Sequence Calibration Wizard
13 to 21
1
Figure 84
on page 127.