5
Maintenance and Repair
Exchanging the CAN Cable
5 The release button [1] (shown in
Step 5 and 9 are for the OLD rear panel design only. See information on
NOTE
the Instant Pilot to an Agilent HPLC Module"
does not have the release button.
6 Unplug the CAN cable from it's connector [2].
7 Plug the new CAN cable into the connector [2].
8 Fit the cable with the cable relief [3] in the front cover.
9 Place the release button into its location [1].
Before you replace the rear panel, observe the release button mechanism at the
NOTE
rear panel. The release button pin must fit into the recess on one end of the
connecting rod.
G4208A User's Guide
the removal. Keep it.
Figure 104 Rear Panel - location of cable, connector and release button
Figure 104
on page 155) may fall out during
on page 16. The new rear panel
"Adding
2
3
155
1