5
Maintenance and Repair
Firmware Updates
Downgrade from B.xx to A.05.1x
1 When the Agilent 1100/1200 series modules are on a firmware platform
2 After boot of the Instant Pilot,
3 Press Cancel (red cross button) on the LAN/RS- 232 Update Tool 2.10 to stop
4 Downgrade all 1100/1200 modules together to the target firmware
5 Close the connection and the LAN/RS- 232 Update Tool 2.10.
G4208A User's Guide
A.06.xx, downgrade the Instant Pilot to A.05.1x first.
During the downgrade process the screen is black.
the Instant Pilot shows "Scanning System ..."
the LAN/RS- 232 Update Tool 2.10 indicates "100% updated" but does not
display the result dialog.
the Update process.
The Instant Pilot is no longer listed in the tree view of the Update Tool.
(A.05.11/13 or A.05.09/10).
When finished,
the LAN/RS- 232 Update Tool 2.10 shows all 1100/1200 modules without
the Instant Pilot.
the Instant Pilot shows all modules of the system in the Welcome screen.