Running Multiple-Vial Analyses
Synchronizing Analyses with External Devices
Synchronizing Analyses with External Devices
With an APG remote connector the system can be connected to external devices
in order to synchronize the analyses. This is necessary when an external device
needs some time in order to get ready for a new analysis and when transmission
of a start request is required (see
further information on interfaces).
The synchronization mode can be chosen in the Configure - System screen.
Figure 86
In any case, do all the analysis preparation using the Instant Pilot.
A "Start" command is used to start the analytical run from the point of injection
NOTE
and is usually issued by the autosampler.
A "Start Request" command causes the autosampler to take the next vial and
place it under the injection needle (see
II)"
The Start button on the Instant Pilot is used to start sequence analysis.
G4208A User's Guide
Choosing the Synchronization mode
on page 101).
"Connecting External Devices"
"APG Remote / Enhanced Remote (Infinity
on page 101 for