Oxi Max Sensor Tabular Event Data - Nellcor OxiMax N-595 Operator's Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

O
M
Sensor Tabular Event Data
XI
AX
N-595
T
O
M
he
sensor tabular event data is a listing of all events recorded on
XI
AX
O
M
the
sensor's memory chip.
XI
AX
The table title shows in the upper left corner. Below the table title is a
six-column table with left-to-right column headings of event number
(#), date (DATE), event start time (START), event duration (DUR),
%SPO
minimum and maximum values during the event (%SPO
2
and pulse rate minimum and maximum values during the event
(BPM).
Event data are listed in chronological order with the most recent event
shown first, at the top of the list, when the tabular Event Summary
display is first accessed. Four events can be displayed
simultaneously; the table must be scrolled to view additional events.
The user can move to the next screen view of the table, the next three
events (the previously displayed bottom or top event is retained as the
fourth event for context when a table is scrolled), using the two
left-most softkeys which are labeled with left- and right-facing arrow
icons, respectively. At the beginning of an event sequence, Event 1
of 5 events, the left-arrow soft key is blank; at the end of a sequence,