Alignment of SITRANS TDL

The emitter and receiver units must be aligned using two adjustment screws. The adjustment
screws align the emitter and receiver units, whereby the laser beam from the emitter unit
reaches the receiver unit, and the receiver unit that faces the emitter unit detects the laser
beam. The receiver/emitter units are mounted on flanges that have a spherical surface. See
flange installation (Page 18) that helps in the alignment procedure.
5.2.5

Alignment procedure

The purpose of alignment is to bring the laser light exiting the emitter to the receiver unit
mounted on the other side of the process site.
The alignment tool contains a light source and an aiming tool which shows the light source
on a target screen. The position of the light source on the target screen corresponds to the
alignment of the system. The aiming tool can be configured for inital (coarse) alignment and
final (fine) alignment.
20
Mount flange on the pro-
cess flange with another
graphite gasket between
flange and purging tube.
Insert adapter with win-
dow into mounted flange.
Check if the clocking bolt
is on the right position to
get into the hole in
adapter.
Tighten the clamp ring (ᴓ
90 mm (3.5 in) on the
adapter rim.
Compact Operating Instructions, 05\2023, A5E51659183-AA – revised 03/2024
ETC- mounting flange on the process flange
Mounting ETC adapter
ETC clamp ring
SITRANS TDL