Setting The Preload - Sartorius Signum 1 Operating Instruction

Complete scales
Hide thumbs Also See for Signum 1:
Table of Contents

Advertisement

Internal Calibration
SIWS models:
Verified scale or scale equipped with
Option E7
In the operating menu under setup:
wp-1
: 1.9., the "Internal calibration"
option (
menu code 1.9.4
) must be set.
setup
wp-1
1. 9.
Calibration,
adjustment
1. 9. 1 * External cal./
adjustment;
default weight
1. 9. 3
External cal./
adjustment;
weight can be
selected under
menu item 1.18. 1)
1. 9. 4
Internal calibra-
tion/adjustment
(only on models
with a built-in
motorized calibra-
tion weight)
1. 9. 8
Set preload
1. 9. 9
Clear preload
J key blocked
1. 9.10
1.10.
Calibration/
adjustment
sequence
1.10. 1
Calibration with
automatic
adjustment
1.10. 2 * Calibration with
adjustment
triggered manually
* = Factory settings
The built-in motorized calibration
weight is applied and removed auto-
matically for internal calibration.
The adjustment procedure is as follows:
§ Select calibration/adjustment:
Press the J key
> The built-in weight is applied
automatically
> The scale is adjusted
> If menu code 1.10.1 is selected
under
Setup
:
wp-1
: 1.10., the scale is
adjusted automatically
> If menu code 1.10.2 is selected under
Setup
:
wp-1
: 1.10., the internal cali-
bration routine can be stopped at this
point, before adjustment is carried out
> The built-in calibration weight is
unloaded from the weighing system
> ISO/GMP-compliant printout:
see page 80

Setting the Preload

Notes on Settings
!The preload can be set only when the
menu access switch is open.
– The "set preload" function must be
assigned to the J key (menu item
1.9.8).
!After setting the preload, close the
menu access switch and reset the J
key to its previous function (e.g., exter-
nal calibration/adjustment with user-
defined weights) under menu item 1.9.
Clearing the Preload
Notes on Settings
!The preload can be cleared only when
the menu access switch is open.
– The "clear preload" function must be
assigned to the J key (menu item
1.9.9).
!After clearing the preload, close the
menu access switch and reset the J
key to its previous function (e.g., exter-
nal calibration/adjustment with user-
defined weights) under menu item 1.9.
43