29.11. Reports from pipette calibration processes

Reports contain data on completed pipette calibration processes. In case of
multichannel pipettes, the reports are generated for each channel separately.
Each report can be previewed, searched by date, exported or printed.
Procedure
Enter submenu <
calibration processes >
Press field of a desired report record, unless it is visible on the top of the
list, wind the list forward using navigating keys
Report's name consists of date and time of its creation,
e.g.: 2012.03.12 11:12:15
Caution:
It is possible to use an option for searching a report.
List of parameters in a report on pipette calibration process:
1. Pipette
2. Serial no.
3. Channel no.
4. Start date
5. End date
6. User
7. Client
8. Number of measurements
9. Operation with ISO 8655
10. Status
11. Temperature [average temperature value during calibration
12. Humidity
13. Pressure
14. Water temperature
15. Z coefficient [value if the Z coefficient for given temperature]

29.12. Series

Database of series contains data saved for a series and measurements of
samples.
Procedure
Enter submenu <
Press <
Databases>, and press <
[name from the database of pipettes]
[serial no. inserted during calibration process]
[name of a user carrying out calibration]
[client name]
[information whether the errors are below permissible
value accepted for the tested volume]
process]
[average humidity value during calibration process]
[average pressure value during calibration process]
Databases>, and press <
Add > key to add a new series to the database
[no. of measurements for each tested
volume]
[information if errors comply with the
standard]
[average water temperature during
calibration process]