Additional Settings Of Part Counting Mode - RADWAG 3Y Series User Manual

Balances 3y series
Table of Contents

Advertisement

16.1. Additional settings of part counting mode

The additional settings enable adjusting the working mode to user's needs and
requirements. To access the setting follow below procedure:
Procedure:
1. Press grey coloured workspace,
2. The display indicates menu of: Settings, Keys, Information, Printout and
Profile
3. Press <Settings> key
4. The display indicates functions related to weighing and parts counting
modes.
Parts counting mode features the following optional functions:
− ACAI, Automatic Accuracy Correction:
YES, mass of a single part is updated during counting process
o
NO, mass of a single part is not updated
o
Means of operation of ACAI function:
1. number of parts (on adding) on balance's weighing pan has to be
greater than has been previously
2. number of parts (on adding) on balance's weighing pan must be less
than twice the amount of which was visible on the display before adding
parts
3. current quantity of parts must be within the ± 0,3 tolerance of the total
value,
4. measurement result has to be stabilized.
− Minimal reference mass: 1unit, 2units, 5units, 10units, it is the
minimal mass value of a single part. Unless the condition is met, counting
process cannot start.
− Result control:
o YES, print and save only those measurements which are included
within the Lo and Hi thresholds
o NO, all measurements are printed and saved
Other function of menu <Settings>:
− Tare mode
− Automatic footer printout
- 67 -

Advertisement

Table of Contents
loading

This manual is also suitable for:

Uya 3yXa 3y.aAs 3yPs 3yApp 3y

Table of Contents