Reports; Formulations Reports - RADWAG PM 10.C32 User Manual

Pm c32 series precision balances
Table of Contents

Advertisement

Values of 3 universal variables
to weighing record.
List of parameters defined for a universal variable:
1. Code (15 characters maximum)
2. Name (43 characters maximum)
3. Value (universal variable value that is to be printed and/or saved to weighing record; 32
characters maximum).
29.

REPORTS

Weighing device software comprises the following reports databases:
• Formulations reports (500 records)
• Weighings (50 000 records)
• Alibi (512 000 records)
FORMULATION, WEIGHINGS and ALIBI REPORTS databases can be imported. Mind that
it is impossible to delete ALIBI database. ALIBI database records can only be exported and
saved to USB flash drive. Names of files with exported data feature serial number and
respective extension (see below table).
Databases

Formulations reports

Weighings
Alibi
Where: xxxxxx – weighing device serial number
Files can be read using ALIBI Reader, PC software designed by RADWAG. You can
download the software from RADWAG website: www.radwag.pl.
Formulations Reports
Formulation reports provide information about carried out formulations. Each report can be
previewed and printed.
Procedure:
Enter <Reports> submenu, select <Formulations reports> parameter.
Select the respective report, if not visible use navigation keys to scroll the reports list
down.
Report name includes date and time, e.g.: 2015.07.12 15:12:15, and report status.
List of formulations reports data:
1. Operator,
2. Formulation,
3. Start date
4. End date
5. Ingredients quantity
6. Measurements quantity
7. Measurements
8. Sum,
9. Target value
10. Difference
,
,
, entered to weighing device memory will be saved
File name and extension
xxxxxx.for
xxxxxx.wei
xxxxxx.ali
- 89 -