28. MASS CONTROL

Working mode <
automatically batched (dosed) onto the balance's weighing pan by an automatic
feeder PA-02/H that is connected to the balance. The working mode
automatically controls a complete batch of tested product after determining its
quantity.

28.1. Working mode activating procedure

Procedure:
Mass control> the software automatically returns to the main
select <
window and displays
Automatically a single on-screen button appears in the bottom bar of the
display:
Start mass control process

28.2. Mass control global settings

Entering the mode settings is carried out by pressing the grey coloured
workspace and selecting an group of parameters named "Settings" <
which contains the following:
Request batch
number
Batch quantity
Low limit
High limit
Print report
Printout

28.3. Mass control process

Before starting the mass control process carry out the following steps:
Mass control> enables controlling mass of a product that is
pictogram in the upper bar,
The function orders requesting a batch
number before starting mass control
process.
Determines quantity of a batch, i.e. number
of product units to be controlled.
A value exceeding which causes accepting a
measurement record by a balance.
A value exceeding which causes omitting a
measurement in the summary, e.g. if two
parts (pills) are simultaneously dispensed on
balance's weighing pan.
A parameter enabling for switching on/off
automatic printout of a report on completing
mass control process.
A parameter enabling selecting whether on
completing the control process the balance
should trigger a standard or a non-standard
printout.