10.3. Balance leveling

Balance Y2 series features AutoLEVEL System, which continuously monitors balance's level
status during its operation. The level status is signaled in the upper right corner of balance's
display. The system continuously controls level, and on detecting level change it indicates
appropriate information on the display and/or signals an alarm. Then the software opens a
window for adjusting the level.
Balance leveling procedure
• Press level status pictogram<
display.
• The display indicates a control panel of leveling function. Balance operator
previews level status, and balance's top elevation.
• Level the balance by turning the adjustable feet in a way indicated by
blinking pictograms on the display < > – the level point should move
towards the center of the level circle.
• As the level point is in the center point, its colour changes from red to
green – which is a confirmation for correct level status.

10.4. Zeroing

In order to zero mass indication press
The mass indication on the display should equal zero, and precise zero
stability
markers should appear.
Zeroing process is tantamount to determining new zero point recognized by
the balance as precise zero. Zeroing is possible only in stable status of display
indication.
Caution:
Zeroing the display indication is possible only within the ±2% range of
instrument's maximal capacity. If zeroed value is above the ±2% of the
maximal capacity, then the software indicates an error message Err2.