Example no. 2 a balance AS 220.3Y:
No.
Tare value
1
220.0000 g
Example no. 3 a balance AS 220.3Y:
No.
Tare value
1
0.0000 g
The user can preview entered data, but data editing is disabled.

29.15. Mass controls

Each product control carried out on a balance is sent to a connected printer and
saved in the < Mass controls >database. Each control saved in the database
features its specific number set at the moment of its completion.
Mass control number format:
y y / M M / d d / H H / m m, where:
yy – year of control end,
MM – month of control end,
dd – day of control end,
HH – hour of control end,
mm – minute of control end.
Balance user can preview data on each control.
Minimum
Means of operation
sample
0.5000 g
The minimum sample refers to all net
weighments that are weighed in a
packaging with optional mass from full
measuring range of a balance (the <TARE>
button applied), as well as on weighing
samples without the packaging (the
<TARE> button is not applied).
Minimum
Means of operation
sample
0.2500 g
The minimum sample refers to all net
weighments that are weighed without a
packaging (the <TARE> button is not
applied).
The software identifies this setting as the
minimum sample is valid only for weighing
samples without a packaging. If tarring
option is applied then the software disables
the pictogram informing on minimum
sample in use, and renders the weighment
as carried out with tare applied, that does
not have a minimum sample criterion
determined.
(mode not available in standard version of the balance)
- 174 -