Transient Withstand Capability

C A U T I O N
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
50 input: 5 Vrms Input protection is enabled in 50
disconnect if greater than 5 Vrms is detected. However the inputs could still be
damaged, depending on the time constant of the signal. The 50
functions when the oscilloscope is powered on.
With 10073C 10:1 probe: CAT I 500 Vpk
With N2871A, N2872A, N2873A 10:1 probe: CAT I 400 Vpk, transient overvoltage
1.25 kVpk
C A U T I O N
±40 V peak CAT I; transient overvoltage 800 Vpk

Environmental Conditions

Environment
Ambient
temperature
Humidity
Altitude
Overvoltage
Category
Pollution Degree
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide

Maximum input voltage at analog inputs

Maximum input voltage at digital channels

Indoor use only.
Operating 0 °C to +55 °C; non-operating –30 °C to +70 °C
Operating: 50% to 95% RH at 40 °C for 5 days.
Non-operating: 90% RH at 65 °C for 24 hr.
Maximum operating altitude: 3,000 m (9,842 ft)
This product is intended to be powered by MAINS that comply to Overvoltage
Category II, which is typical of cord-and-plug connected equipment.
The InfiniiVision 4000 X-Series oscilloscopes may be operated in
environments of Pollution Degree 2 (or Pollution Degree 1).
Reference
mode and the 50
load will
input protection only
22
367