9
Characteristics and Specifications

6000A Series and 6000L Series Transient Withstand Capability

C A U T I O N
C A U T I O N
C A U T I O N
C A U T I O N
344
Maximum input voltage for analog inputs:
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
with 10073C or 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
Do not exceed 5 Vrms in 50 Ω mode on the 2-channel models. Input
protection is enabled in 50 Ω mode, and the 50 Ω load will disconnect if
greater than 5 Vrms is detected. However, the input could still be
damaged, depending on the time constant of the signal.
The 50 Ω input protection mode only functions when the oscilloscope
is powered on.
Maximum input voltage for logic channels:
±40 V peak CAT I; transient overvoltage 800 Vpk