Dual-Channel (N2820A Probe)
Measurement
DC RMS
AC RMS
When using the N2820A probe to make measurements on a
battery- powered (floating) device, always connect the supplied ground lead
between ground on your device and the probe's ground connector as
shown in the following figure. Simply snap the end of the ground lead
onto the probe's connector. Without the ground connection, the probe's
common- mode input amplifier cannot properly display waveforms.
Figure 46

Snapshot All

The Snapshot All measurement type displays a popup containing a
snapshot of all the single waveform measurements.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
Measurements on Battery Powered Devices Using the N2820A Probe
Notes
See
"DC RMS"
on page 251.
See
"AC RMS"
on page 252.