I2S triggering and serial decode requires Option AUDIO or the
DSOX4AUDIO upgrade.
Only one I2S serial bus can be decoded at a time.
N O T E
Setup for I2S Signals
I
of connecting the oscilloscope to the serial clock, word select, and serial
data lines and then specifying the input signal threshold voltage levels.
To set up the oscilloscope to capture I2S signals:
1 Press [Label] to turn on labels.
2 Press [Serial].
3 Press the Serial softkey, turn the Entry knob to select the desired slot
4 Press the Mode softkey; then, select I2S trigger type.
5 Press the Signals softkey to open the I
Agilent InfiniiVision 4000 X-Series Oscilloscopes
User's Guide
26
I2S Triggering and Serial Decode
Setup for I2S Signals
I2S Triggering
I2S Serial Decode
2
S (Inter- IC Sound or Integrated Interchip Sound) signals setup consists
(Serial 1 or Serial 2), and press the softkey again to enable decode.