Transient Withstand Capability

C A U T I O N
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
With N2862A or N2863A 10:1 probe: 300 Vrms
C A U T I O N
±40 V peak CAT I; transient overvoltage 800 Vpk

Environmental Conditions

Environment
Ambient
temperature
Humidity
Altitude
Overvoltage
Category
Pollution Degree
Pollution Degree
Definitions
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

Maximum input voltage at analog inputs

Maximum input voltage at digital channels

Indoor use only.
Operating 0 °C to +55 °C; non-operating –40 °C to +71 °C
Operating: Up to 80% RH at or below +40 °C. Up to 45% RH up to +50 °C.
Non-operating: Up to 95% RH up to +40 °C. Up to 45% RH up to +50 °C.
Operating and non-operating to 4,000 m (13,123 ft)
This product is intended to be powered by MAINS that comply to Overvoltage
Category II, which is typical of cord-and-plug connected equipment.
The InfiniiVision 2000/3000 X-Series oscilloscopes may be operated in
environments of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
The pollution has no influence. Example: A clean room or climate controlled
office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive
pollution occurs which becomes conductive due to condensation which is
expected. Example: Sheltered outdoor environment.