To Recall Setup Files; To Recall Mask Files; To Recall Reference Waveform Files From A Usb Storage Device - Agilent Technologies InfiniiVision 4000 X-Series User Manual

Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

18
Save/Email/Recall (Setups, Screens, Data)
You can also configure the [Quick Action] key to recall setups, masks, or
reference waveforms. See
page 345.

To recall setup files

Setup files can be recalled from one of 10 internal (\Agilent Flash)
locations or from an external USB storage device.
1 Press [Save/Recall] > Recall > Recall:; then, turn the Entry knob to select
2 Press the softkey in the second position and use the Entry knob to
3 Press the Press to Recall softkey.
4 If you would like to clear the display, press Clear Display.

To recall mask files

Mask files can be recalled from one of four internal (\Agilent Flash)
locations or from an external USB storage device.
1 Press [Save/Recall] > Recall > Recall:; then, turn the Entry knob to select
2 Press the softkey in the second position and use the Entry knob to
3 Press the Press to Recall softkey.
4 If you would like to clear the display or clear the recalled mask, press

To recall reference waveform files from a USB storage device

1 Press the [Save/Recall] key.
2 In the Save/Recall Menu, press the Recall softkey.
318
Setup (*.scp).
navigate to the file to recall. See
page 315.
A message indicating whether the recall was successful is displayed.
Mask (*.msk).
navigate to the file to recall. See
page 315.
A message indicating whether the recall was successful is displayed.
Clear Display or Clear Mask.
"Configuring the [Quick Action] Key"
"To navigate storage locations"
"To navigate storage locations"
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide