Agilent Technologies 33220A User Manual page 311

20 mhz waveform generator
Table of Contents

Advertisement

786
787
788
800
810
5
Chapter 5 Error Messages
Arbitrary Waveform Errors
Not able to delete a built-in arb waveform
You cannot delete any of the five built-in waveforms: "EXP_RISE",
"EXP_FALL", "NEG_RAMP", "SINC", and "CARDIAC".
Not able to delete the currently selected active arb waveform
You cannot delete the arbitrary waveform that is currently being output
(FUNC:USER command).
Cannot copy to VOLATILE arb waveform
The DATA:COPY command copies the waveform from volatile memory to
the specified name in non-volatile memory. The source for the copy is
always "VOLATILE". You cannot copy from any other source and you
cannot copy to "VOLATILE".
Block length must be even
The function generator represents binary data as 16-bit integers,
which are sent as two bytes (DATA:DAC VOLATILE command).
State has not been stored
The storage location that you specified in the *RCL command was not
used in a previous *SAV command. You cannot recall the instrument
state from a storage location that is empty.