Device Dependent Errors

-313
Calibration memory lost;
memory corruption detected
The non-volatile memory used to store the function generator's calibration
constants has detected a checksum error. This error may be the result of
a device failure or extreme conditions such as lightning or strong
magnetic fields.
Save/recall memory lost;
-314
memory corruption detected
The non-volatile memory used to store instrument states has detected a
checksum error. This error may be the result of a device failure or
extreme conditions such as lightning or strong magnetic fields.
-315
Configuration memory lost;
memory corruption detected
The non-volatile memory used to store the function generator's
configuration settings (i.e., remote interface settings) has detected a
checksum error. This error may be the result of a device failure or
extreme conditions such as lightning or strong magnetic fields.
-350
Queue overflow
The error queue is full because more than 20 errors have occurred.
No additional errors are stored until you remove errors from the queue.
The error queue is cleared by the *CLS (clear status) command or when
the power is cycled. The errors are also cleared when you read the queue.
The error queue is not cleared by an instrument reset (*RST command).
Chapter 5 Error Messages
Device Dependent Errors
4
5
303