Sweep Trigger Source - Agilent Technologies 33220A User Manual

20 mhz waveform generator
Table of Contents

Advertisement

3
Chapter 3 Features and Functions
Frequency Sweep

Sweep Trigger Source

In the sweep mode, the function generator outputs a single sweep when a
trigger signal is received. After one sweep from the start frequency to the
stop frequency, the function generator waits for the next trigger while
outputting the start frequency.
• Sweep trigger source: Internal, External, or Manual. The default
is Internal.
• When the Internal (immediate) source is selected, the function
generator outputs a continuous sweep at a rate determined by the
sweep time specified.
• When the External source is selected, the function generator will
accept a hardware trigger applied to the rear-panel Trig In connector.
The function generator initiates one sweep each time Trig In receives
a TTL pulse with the specified polarity.
• The trigger period must be greater than or equal to the specified
sweep time plus 1 ms.
• When the Manual source is selected, the function generator outputs
one sweep each time the front-panel
• Front-Panel Operation: Press the Trigger Setup softkey and then
select the desired source by pressing the Source softkey.
To specify whether the function generator triggers on the rising or
falling edge of the Trig In connector, press the Trigger Setup softkey.
Then select the desired edge by pressing the Slope softkey.
• Remote Interface Operation:
TRIGger:SOURce {IMMediate|EXTernal|BUS}
Use the following command to specify whether the function generator
triggers on the rising or falling edge of the signal on the Trig In
connector.
TRIGger:SLOPe {POSitive|NEGative}
See "Triggering" on page 115 for more information.
104
key is pressed.

Advertisement

Table of Contents
loading

Table of Contents