-221
Settings conflict;
FSK turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.
-221
Settings conflict;
FM turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.
-221
Settings conflict;
AM turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.
Settings conflict;
-221
PM turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.
-221
Settings conflict;
PWM turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.
-221
Settings conflict;
sweep turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.
Settings conflict;
-221
not able to modulate this function
The function generator cannot generate an AM, FM, PM, or FSK
modulated waveform using the pulse, noise, or dc voltage functions.
Chapter 5 Error Messages