The Status Byte condition register is cleared when:
• You execute the *CLS (clear status) command.
• You read the event register from one of the other register groups
(only the corresponding bits are cleared in the condition register).
The Status Byte enable register is cleared when:
• You execute the *SRE 0 command.
• You turn the power on and have previously configured the function
generator to clear the enable register using the *PSC 1 command.
Note that the enable register will not be cleared at power-on if you
have configured the function generator using the *PSC 0 command.
Chapter 4 Remote Interface Reference
The SCPI Status System
4
263