Advertisement

4. Leaving the phase setting at its default (zero) on the first 33220A, use
the Adjust Phase feature to adjust the phase of the second 33220A to
align the output signals:
Out of Phase
You can use the Set 0 Phase feature to set a new zero-phase reference
point with the two function generators now aligned.
• Front-Panel Operation: Press
softkey, then Adjust Phase. Set the phase angle (±360 degrees) using
the knob and cursor keys or the numeric keypad. Once you have
aligned the signals on the scope, press Set 0 Phase to set a new zero-
phase reference point.
• Remote Interface Operation: Use the following command to set the
phase offset (the query returns the currently set offset):
PHASe {< angle >|MINimum|MAXimum}
PHASe? [MINimum|MAXimum]
• The following command sets a new zero-phase reference point:
PHASe:REFerence
Application Note 1426: "How to Connect Two or More Signal
Generators to Create a Multi-Channel Waveform Generator" provides
additional information on this subject.