Chapter 4 Remote Interface Reference
Using the APPLy Command
APPLy:SINusoid [<frequency> [,<amplitude> [,<offset>] ]]
Output a sine wave with the specified frequency, amplitude, and
dc offset. The waveform is output as soon as the command is executed.
APPLy:SQUare [<frequency> [,<amplitude> [,<offset>] ]]
Output a square wave with the specified frequency, amplitude, and
dc offset. This command overrides the current duty cycle setting and
automatically selects 50%. The waveform is output as soon as the
command is executed.
APPLy:RAMP [<frequency> [,<amplitude> [,<offset>] ]]
Output a ramp wave with the specified frequency, amplitude, and
dc offset. This command overrides the current symmetry setting and
automatically selects 100%. The waveform is output as soon as the
command is executed.
APPLy:PULSe [<frequency> [,<amplitude> [,<offset>] ]]
Output a pulse wave with the specified frequency, amplitude, and
dc offset. The waveform is output as soon as the command is executed.
• This command preserves the current pulse width setting
(FUNC:PULS:WIDT command) or pulse duty cycle setting
(FUNC:PULS:DCYC command), depending on which has been selected
to "hold" (FUNC:PULS:HOLD command). The edge time setting
(FUNC:PULS:TRAN command) is also preserved. However, based on
the specified frequency, the function generator will adjust the pulse
width or edge time to comply with the frequency restrictions for pulse
waveforms. See page 192 for more information on setting the pulse
width and edge time.