Agilent Technologies 33220A User Manual page 288

20 mhz waveform generator
Table of Contents

Advertisement

-158
String data not allowed
A character string was received but is not allowed for this command.
Check the list of parameters to verify that you have used a valid
parameter type. Example: BURS:NCYC 'TEN'
-161
Invalid block data
Applies only to the DATA:DAC VOLATILE command.
For a definite-length block, the number of bytes of data sent does not
match the number of bytes that you specified in the block header. For an
indefinite-length block, an EOI (End-or-Identify) was received without
an accompanying <new line> character.
-168
Block data not allowed
Data was sent to the function generator in arbitrary block format but
this command does not accept this format. Check to see if you have sent
the correct data type with the command. Example: BURS:NCYC #10
-170 to -178
Expression errors
The function generator does not accept mathematical expressions.
Chapter 5 Error Messages
Command Errors
4
5
287