Chapter 5 Error Messages
Execution Errors
Settings conflict;
PWM only available in pulse function
The function generator cannot generate a PWM modulated waveform
using any function other than pulse.
Settings conflict;
not able to sweep this function
The function generator cannot generate a sweep using the pulse, noise,
or dc voltage functions.
Settings conflict;
not able to burst this function
The function generator cannot generate a burst using the dc voltage
function.
Settings conflict;
not able to modulate noise, modulation turned off
The function generator cannot generate a modulated waveform using the
noise function. The selected modulation mode has been turned off.
Settings conflict;
not able to sweep pulse, sweep turned off
The function generator cannot generate a sweep using the pulse
function. The sweep mode has been turned off.
Settings conflict;
not able to modulate dc, modulation turned off
The function generator cannot generate a modulated waveform using the
dc voltage function. The selected modulation mode has been turned off.
Settings conflict;
not able to sweep dc, sweep turned off
The function generator cannot generate a sweep using the dc voltage
function. The sweep mode has been turned off.
Settings conflict;
not able to burst dc, burst turned off
The function generator cannot generate a burst using the dc voltage
function. The burst mode has been turned off.