Agilent Technologies 33220A User Manual page 218

20 mhz waveform generator
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference
Frequency Sweep Commands
Sweep Commands
FREQuency:STARt {<frequency>|MINimum|MAXimum}
FREQuency:STARt? [MINimum|MAXimum]
Set the start frequency (used in conjunction with the stop frequency).
Select from 1 µHz to 20 MHz (limited to 200 kHz for ramps and 6 MHz
for arbitrary waveforms). The default is 100 Hz. MIN = 1 µHz. MAX =
20 MHz. The :STAR? query returns the start frequency in hertz.
• To sweep up in frequency, set the start frequency < stop frequency.
To sweep down in frequency, set the start frequency > stop frequency.
FREQuency:STOP {<frequency>|MINimum|MAXimum}
FREQuency:STOP? [MINimum|MAXimum]
Set the stop frequency (used in conjunction with the start frequency).
Select from 1 µHz to 20 MHz (limited to 200 kHz for ramps and 6 MHz
for arbitrary waveforms). The default is 1 kHz. MIN = 1 µHz. MAX =
20 MHz. The :STOP? query returns the stop frequency in hertz.
4
217

Advertisement

Table of Contents
loading

Table of Contents