Chapter 5 Error Messages
Execution Errors
Data out of range;
amplitude; value clipped to ...
This generic message indicates that the waveform amplitude has been
limited to an upper or lower boundary.
Data out of range;
offset; value clipped to ...
This generic message indicates that the offset voltage has been limited to
an upper or lower boundary.
Data out of range;
frequency in burst mode; value clipped to ...
This generic message indicates that the frequency has been limited to an
upper or lower boundary as dictated by the burst period.
Data out of range;
frequency in FM; value clipped to ...
This generic message indicates that the carrier frequency has been
limited to a lower boundary as determined by the FM:DEV command.
The carrier frequency must always be greater than or equal to the
frequency deviation.
Data out of range;
marker confined to sweep span; value clipped to ...
This generic message indicates that the specified marker frequency is
outside the range of the start frequency and stop frequency. The marker
frequency must be between the specified start frequency and stop
frequency. If you attempt to set the marker frequency to a frequency
not in this range, the function generator will automatically set the
marker frequency equal to the start frequency or stop frequency
(whichever is closer). This error is generated only when both the sweep
mode and the marker frequency are enabled.
Data out of range;
pulse width; value clipped to ...
This generic message indicates that the desired pulse width is limited to
an upper or lower boundary which is dictated by the instrument
hardware.