Agilent Technologies 33220A User Manual page 172

20 mhz waveform generator
Table of Contents

Advertisement

Output Frequency
• For the frequency parameter of the APPLy command, the output
frequency range depends on the function specified. You can substitute
"MINimum", "MAXimum", or "DEFault" in place of a specific value
for the frequency parameter. MIN selects the lowest frequency
allowed for the function specified and MAX selects the highest
frequency allowed. The default frequency is 1 kHz for all functions.
Function
Minimum Frequency
Sine
Square
Ramp
Pulse
Noise, DC
Not Applicable
Arbs
• Limits Due to Function: The frequency limits are determined by the
function specified in the APPLy command. The APPLy command
always sets both the function and the frequency, and the specified
frequency must be appropriate for the function. For example, the
command APPL:RAMP 20 MHz results in a "Data out of range" error.
The frequency is set to 200 kHz, which is the maximum for a ramp
waveform.
Chapter 4 Remote Interface Reference
Using the APPLy Command
Maximum Frequency
1 µHz
1 µHz
1 µHz
500 µHz
Not Applicable
1 µHz
20 MHz
20 MHz
200 kHz
5 MHz
6 MHz