Marker Frequency

If desired, you can set the frequency at which the signal on the front-
panel Sync connector goes to a logic low during the sweep. The Sync
signal always goes from low to high at the beginning of the sweep.
• Marker frequency: 1 µHz to 20 MHz (limited to 200 kHz for ramps
and 6 MHz for arbitrary waveforms). The default is 500 Hz.
• When the sweep mode is enabled, the marker frequency must be
between the specified start frequency and stop frequency. If you
attempt to set the marker frequency to a frequency not in this range,
the function generator will automatically set the marker frequency
equal to the start frequency or stop frequency (whichever is closer).
• The Sync enable setting is overridden by enabling the marker used
with the sweep mode (see page 68). Therefore, when the marker is
enabled (and the sweep mode is also enabled), the Sync setting is
ignored.
• Front-Panel Operation: After enabling sweeps, press the Marker
softkey. Then use the knob or numeric keypad to enter the desired
marker frequency.
• Remote Interface Operation:
MARKer:FREQuency {<frequency>|MINimum|MAXimum}
MARKer {Off|On}
Chapter 3 Features and Functions
Frequency Sweep
3