Chapter 5 Error Messages
Execution Errors
Settings conflict;
trigger output connector used by trigger external
When the external trigger source is selected (TRIG:SOUR EXT command),
the function generator automatically disables the "trigger out" signal.
The rear-panel Trig connector cannot be used for both operations at the
same time.
Settings conflict;
frequency reduced for user function
For arbitrary waveforms, the output frequency is limited to 6 MHz.
When you change to the arbitrary waveform function (APPL:USER or
FUNC:USER command) from a function that allows a higher frequency,
the function generator will automatically adjust the frequency to 6 MHz.
Settings conflict;
frequency changed for pulse function
For pulse waveforms, the output frequency is limited to 5 MHz.
When you change to the pulse function (APPL:PULS or FUNC:PULS
command) from a function that allows a higher frequency, the function
generator will automatically adjust the frequency to 5 MHz.
Settings conflict;
frequency reduced for ramp function
For ramp waveforms, the output frequency is limited to 200 kHz.
When you change to the ramp function (APPL:RAMP or FUNC:RAMP
command) from a function that allows a higher frequency, the function
generator will automatically adjust the frequency to 200 kHz.
Settings conflict;
frequency made compatible with burst mode
For an internally-triggered burst, the output frequency is limited to a
minimum of 2.001 mHz. The function generator has adjusted the
frequency to be compatible with the current settings.
Settings conflict;
burst turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.