9 Define the next waveform point.
Press the Point # softkey and then turn the knob to move to Point #2.
Press the Time softkey to set the time for the current point (this softkey
is not available for Point #1). Press the Voltage softkey to set the voltage
level for the current point.
For this example, set the time to 2 ms and the voltage level to 3.0 V.
10 Define the remaining waveform points.
Using the Time and Voltage softkeys, define the remaining waveform
points using the values shown in the table below.
Point
Time Value
1
0 s
2
2 ms
3
4 ms
4
7 ms
• The time value of the last point that can be defined in the waveform
must be less than the specified cycle period.
• The waveform editor automatically connects the last waveform point
to the voltage level of first point to create a continuous waveform.
• To insert additional points after the current waveform point, press
the Insert Point softkey. The new point is inserted midway between
the current point and the next defined point.
• To remove the current waveform point, press the Remove Point
softkey. The remaining points are joined using the interpolation
method currently selected. You cannot remove Point #1 since the
waveform must have a defined initial value.
Chapter 3 Features and Functions
Voltage Value
0 V
3 V
1 V
0 V
Arbitrary Waveforms
123
3