-221
Settings conflict;
high level changed due to low level
You can set the levels to a positive or negative value, but note that the
high level must always be greater than the low level. If you specify a low
level that is greater than the existing high level, the function generator
will automatically set the high level to be 1 mV greater than the low
level.
-222
Data out of range;
value clipped to upper limit
The specified parameter is outside of the function generator's capability.
The function generator has adjusted the parameter to the maximum
value allowed. Example: PHAS 1000
-222
Data out of range;
value clipped to lower limit
The specified parameter is outside of the function generator's capability.
The function generator has adjusted the parameter to the minimum
value allowed. Example: PHAS -1000
Data out of range;
-222
pulse edge time limited by period; value clipped to upper limit
The specified edge time must fit within the existing period and width.
The function generator will limit the edge time as needed to
accommodate the existing period.
-222
Data out of range;
pulse width limited by period; value clipped to ...
The specified pulse width must be less than the difference between the
period and the edge time as shown below. The function generator will
adjust the pulse width as needed to accommodate the specified period.
Pulse Width < Period – (1.6 X Edge Time)
-222
Data out of range;
pulse duty cycle limited by period; value clipped to ...
The specified pulse duty cycle must conform to the following period and
edge time restriction. The function generator will adjust the pulse duty
cycle as needed to accommodate the specified period.
Duty Cycle < 100 X (Period – (1.6 X Edge Time)) ÷ Period
Chapter 5 Error Messages
Execution Errors
4
5
297