SYSTem:COMMunicate:LAN:TELNet:PROMpt < string >
SYSTem:COMMunicate:LAN:TELNet:PROMpt?
Sets the command prompt seen when using a Telnet session to
communicate with the 33220A. The default prompt is "33220A>".
• The 33220A uses LAN port 5024 for Telnet sessions.
• Telnet session can typically be started as follows from a host
computer shell:
telnet <IP_address> <port>
For example:
telnet 169.254.2.20 5024
To exit a Telnet session, press <Cntr-D>.
• The prompt is a quoted string of up to 15 characters.
• The command prompt is stored in non-volatile memory, and does not
change when power has been off, or after a instrument reset (*RST).
SYSTem:COMMunicate:LAN:TELNet:WMESsage < string >
SYSTem:COMMunicate:LAN:TELNet:WMESsage?
Sets the welcome message seen when using a Telnet session to
communicate with the 33220A. The default welcome message is
"Welcome to Agilent's 33220A Waveform Generator".
• The 33220A uses LAN port 5024 for Telnet sessions.
• The message is a quoted string of up to 63 characters.
• The welcome message is stored in non-volatile memory, and does not
change when power has been off, or after a instrument reset (*RST).

LAN Port Usage The Agilent 33220A uses LAN port 5024 for Telnet

sessions, but uses LAN port 5025 for Socket sessions. For further
information on socket sessions, see the white paper "Socket Connections
for LAN-enabled Instruments" found on the Agilent 33220A Product
Reference CD.