Agilent Technologies 33220A User Manual page 295

20 mhz waveform generator
Table of Contents

Advertisement

-221
-221
-221
5
-221
Chapter 5 Error Messages
Execution Errors
Settings conflict;
edge time decreased due to pulse width
For a pulse waveform, the function generator will automatically adjust
the waveform parameters in the following order as needed to generate a
valid pulse: (1) edge time, (2) pulse width or duty cycle, and then (3)
period.
In this case, the function generator has decreased the edge time to
accommodate the specified pulse width.
Edge Time < 0.625 X Pulse Width
Settings conflict;
edge time decreased due to pulse duty cycle
For a pulse waveform, the function generator will automatically adjust
the waveform parameters in the following order as needed to generate a
valid pulse: (1) edge time, (2) pulse width or duty cycle, and then (3)
period.
In this case, the function generator has decreased the edge time to
accommodate the specified pulse duty cycle.
Edge Time < 0.625 X Period X Duty Cycle ÷ 100
Settings conflict;
amplitude changed due to function
In some cases, the amplitude limits are determined by the output units
currently selected. This may occur when the units are Vrms or dBm due
to the differences in crest factor for the various output functions.
For example, if you output a 5 Vrms square wave (into 50 ohms) and
then change to the sine wave function, the function generator will
automatically adjust the output amplitude to 3.536 Vrms (upper limit for
sine waves in Vrms).
Settings conflict;
offset changed on exit from dc function
In the dc voltage function, the voltage level is controlled by adjusting the
offset voltage (the current amplitude is ignored). When you select a
different function, the function generator will adjust the offset voltage as
needed to be compatible with the current amplitude setting.
294

Advertisement

Table of Contents
loading

Table of Contents