3.1.3 Dual Range Calibration

If the analyzer is being operated in Dual Range mode or Auto-Ranging mode, then the High and
Low ranges must be independently calibrated. When the analyzer is in Dual or Auto Range mode
you will be prompted to enter the range to calibrate whenever you enter a calibration command
from the front panel. Press HIGH or LOW followed by ENTR button to proceed with the
calibration. To calibrate the other range you must exit to the sample menu and restart the
calibration. See Section 4.6 for more information on the Range Modes. The following procedure
shows an example of how to calibrate the two ranges with calibration gas coming in through the
sample port:
Step
Action
1.
Press CAL
2.
Press LOW-ENTR
3.
Wait 15 min.
4.
Press ZERO-ENTR
5.
Press CONC
6.
Key in span concentration
7.
Press ENTR
8.
Wait 15 min.
9.
Press SPAN-ENTR
10.
Press EXIT
Repeat steps 1-10 for High range.
Comment
Analyzer enters M-P calibration mode. Calibration gas
source should be set to deliver zero gas to the sample
port.
Select range to calibrate.
Wait for CO reading to stabilize at zero value.
Changes calibration equations for Low range so
analyzer will read zero.
Enter span gas concentration for Low range.
Enter span gas concentration for Low Range. Set
calibration gas source to deliver span concentration.
Wait for CO reading to stabilize at span value.
Changes calibration equations for Low range so
analyzer will read span value.
Exits back to sample menu.